Abstract

Concentration polarization and secondary membrane formation change the membrane hydraulic permeability of capillary filters during hemofiltration reducing initial value filtration rates up to 50%. This leads to a significant loss of filter efficiency which must be taken into consideration when designing filters for long-term application such as in implantable artificial kidneys. By measuring blood density uninterrupted over a period of time using the mechanical oscillatory technique it is possible to follow the dynamics of protein deposition at the interface between the blood and capillary walls. The resulting picture of the deposition behaviour can lead to a better understanding of time-dependent filtration where flow and pressure conditions change. Protein deposition (Pt) with polysulfone membranes in relation to the effective capillary surface was in the range of 2.2. mg/(h x cm2) according to a logarithmic function (Pt = a + b x log(t)).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.