Abstract

The space magnetic field is produced by a current-carrying coil. According to the triangle rule, whenever the difference of results measured by the magnetometer between the coil switching the current on and off is less than the calculated value of the coil magnetic field, the magnetometer functional mode is normal. When the magnetometer is at the critical state of the normal work, the coil magnetic field gradient is its gradient tolerance. The Cs-3 optical-pumping magnetometer gradient tolerance is 20000nT/m by experiment. Measurement result can be used for magnetic environment optimization and cesium probe assembly.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.