Abstract

A measurement technique enabling the determination of low uranium concentration utilizing system used for time resolved laser induced fluorescence spectroscopy (TRLFS), based on a commercially available kinetic phosphorescence analysis method has been developed. The effect of the matrix, complexing agent, and the level of acidification on the fluorescence spectra and integral intensity was studied. The method is highly influenced by the sample matrix; particularly if different starting uranium salts are in the calibration solutions and in the samples. In the initial study of these effects, TRLFS measurements have been used in order to clarify this behavior.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.