Abstract

The aim of this study was to assess the conductive loss component (CLC) by brainstem electric response audiometry. A bone-conducted noise was used to mask out the response to a conventional air-conducted click stimulus. The difference between the levels of the click and the noise is defined as the masked threshold to noise ratio (MTNR). This MTNR was determined for 21 normal ears (MTNR = -13 +/- 5 dB). The increase in MTNR compared to this normative value is a measure of the CLC. For 10 ears with an artificially induced purely conductive loss, the increase in MTNR is in good agreement with the results of conventional pure-tone and brainstem electric response audiometry.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.