Abstract

PurposeWe further optimized and evaluated the safety of the magnetic levator prosthesis (MLP) for temporary management of severe blepharoptosis, and compared efficacy and comfort against the ptosis crutch.MethodsThe interpalpebral fissure (IPF) of participants (n = 12) with ptosis was measured during attempted eyelid opening, volitional closing, and spontaneous closing with no device, ptosis crutch, or the MLP. A 10-point scale documented comfort. Additionally, a 20 minute and then 1 week trial of the MLP was offered. Safety measures were skin erythema rating, change in visual acuity, and change in corneal staining.ResultsThe MLP and crutch opened the eye (IPF 11.2 and 9.3 mm), but the MLP allowed better volitional closure (IPF 1.0 vs. 4.9 mm, P = 0.009), but was no better in allowing spontaneous blink (IPF 7.5 vs. 7.7 mm, P = 0.722). Both devices were equally comfortable (both median 8/10 comfort, P = 0.46). With extended use, opening with the MLP showed IPF 9.24 mm at 20 minutes and 9.46 mm at 1 week, and volitional closure was IPF 0.95 and 0.52 mm, respectively. Closure on spontaneous blink improved with extended wear to IPF 5.14 and 5.18 mm, respectively (P = 0.002). Two participants exhibited moderate skin erythema and one had increased corneal staining without change in acuity.ConclusionsThe MLP is safe and feasible for temporary correction of severe ptosis.Translational RelevanceFirst group data in patients showing successful reanimation of the eyelid with magnetic force.