Abstract
LUCI1 and LUCI2 are a pair multi-mode, fully cryogenic near-infrared instruments installed at the Large Binoc- ular Telescope (LBT). The instruments provide imaging, long-slit and multi-object spectroscopy over a 4/ FoV in seeing-limited mode. Ground-layer AO (GLAO) correction for imaging and spectroscopy over the 4/ FoV is available using the ARGOS laser system, as well as diffraction-limited AO over a 30// FoV using the LBT first light AO (FLAO) system with natural guide stars. Internal flexure of the instrument is taken care of by passive and active flexure compensation. Image shifts in seeing-limited modes are compensated by a passive flexure con- trol algorithm using pre-defined look-up tables. For AO observations, passive compensation is replaced by active control. In the following, we present the details of the newly developed active flexure compensation algorithm for the LUCI instruments. We also describe some hardware modifications to the instruments and the results obtained with active flexure compensation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.