Abstract

The feasibility of implementing magnetic struts into drug-eluting stents (DESs) to mitigate the adverse hemodynamics which precipitate stent thrombosis is examined. These adverse hemodynamics include platelet-activating high wall shear stresses (WSS) and endothelial dysfunction-inducing low wall shear stresses. By magnetizing the stent struts, two forces are induced on the surrounding blood: (1) magnetization forces which reorient red blood cells to align with the magnetic field and (2) Lorentz forces which oppose the motion of the conducting fluid. The aim of this study was to investigate whether these forces can be used to locally alter blood flow in a manner that alleviates the thrombogenicity of stented vessels. Two-dimensional steady-state computational fluid dynamics (CFD) simulations were used to numerically model blood flow over a single magnetic drug-eluting stent strut with a square cross section. The effects of magnet orientation and magnetic flux density on the hemodynamics of the stented vessel were elucidated in vessels transporting oxygenated and deoxygenated blood. The simulations are compared in terms of the size of separated flow regions. The results indicate that unrealistically strong magnets would be required to achieve even modest hemodynamic improvements and that the magnetic strut concept is ill-suited to mitigate stent thrombosis.