Abstract

IntroductionAlthough serum-providing blood tubes with a barrier are still widely used due to their significant advantages, the use of blood tubes with a barrier to provide plasma is becoming widespread. We compared 22 analytes in a BD Vacutainer® Barricor LH Plasma tube for local clinical validation of this new lithium heparin tube with a barrier.Materials and methodsSamples from 44 volunteers were collected in different tubes (Becton Dickinson and Company): Z tube without additive (reference), clot-activator tube with gel (SST), lithium heparin tube without gel (LiH), and lithium heparin tube with barrier (Barricor). Analyte concentrations in different tubes were compared with the reference tube. All tubes were also evaluated according to additional testing (different centrifugation durations, blood-sampling techniques and individual differences).ResultsAspartate aminotransferase (AST), glucose (Glc), potassium (K), lactate dehydrogenase (LD), sodium (Na), and total protein (TP) had a significant bias in Barricor (9.19%, - 3.24%, - 4.88%, 21.60%, - 0.40%, 5.03%, respectively) relative to the reference tube. There was no statistical difference between different centrifugation durations and individual differences for AST, K and LD in LiH and/or Barricor (P > 0.05). There was a significant bias for LD between LiH and Barricor in terms of blood-sampling techniques (21.2% and 12.4%, respectively).ConclusionsRecently, the use of plasma has become prominent due to some of its advantages. In this study, plasma AST, K, LD, Glc and TP levels in Barricor were clinically different in comparison to serum. The results of additional tests showed that higher levels of LD in Barricor did not result from haemolysis, and they might be related to other factors including number of platelets, cellular fragility, or functional environment.

Highlights

  • Serum-providing blood tubes with a barrier are still widely used due to their significant advantages, the use of blood tubes with a barrier to provide plasma is becoming widespread

  • Plasma AST, K, lactate dehydrogenase (LD), Glc and total protein (TP) levels in Barricor were clinically different in comparison to serum

  • The results of additional tests showed that higher levels of LD in Barricor did not result from haemolysis, and they might be related to other factors including number of platelets, cellular fragility, or functional environment

Read more

Summary

Introduction

Serum-providing blood tubes with a barrier are still widely used due to their significant advantages, the use of blood tubes with a barrier to provide plasma is becoming widespread. Other important advantages that make the use of these tubes so widespread are the lower need for aliquot tubes, improved analytical stability, and reduced haemolysis rates by facilitating rapid separation of serum from the cellular constituents of blood [1]. In proportion to serum, 15 – 20% more plasma can be obtained from whole blood, and the risk of thrombolysis and haemolysis in plasma is lower than in serum This means that because the blood cells remain intact in the healthy donors’ plasma, no pseudohyperkalemia occurs in plasma in contrast to serum [2].

Objectives
Methods
Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.