Abstract
In conscious subjects breathing spontaneously the performance of the Lack semi-closed system resembled the Magill attachment in that rebreathing was only detected at fresh gas flows of less than resting minute volume. With the Bain semi-closed system rebreathing was detectable at fresh gas flows of the order of 2.5 times the minute volume. The two co-axial circuits behaved similarly for anaesthetised patients. In adults breathing spontaneously the Lack system is efficient and more economical than the Bain.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.