Abstract

In conscious subjects breathing spontaneously the performance of the Lack semi-closed system resembled the Magill attachment in that rebreathing was only detected at fresh gas flows of less than resting minute volume. With the Bain semi-closed system rebreathing was detectable at fresh gas flows of the order of 2.5 times the minute volume. The two co-axial circuits behaved similarly for anaesthetised patients. In adults breathing spontaneously the Lack system is efficient and more economical than the Bain.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.