Abstract

This paper provides a technical description of the neutron emission profile monitor as used routinely at the Joint European Torus (JET), and includes representative examples of its operational capabilities. The primary function of this instrument is to measure the neutron emission as a function of both position and time in a poloidal (vertical along major radius) section through the torus. For the first time the spatially localised effects of sawteeth (magnetic relaxation phenomena) have been observed using a neutron diagnostic. The total (global) neutron emission can be obtained from the profile monitor data by performing a volume integral over the plasma; the absolute neutron emission rates agree with those obtained from the JET time-resolved neutron monitor to within ±15%. This was the first such instrument routinely in use on any tokamak. It provides unique data which are independent of all other diagnostic measurements.