Abstract

The ion circus is a miniature storage ring formed by a segmented radiofrequency mass filter bent into a circle. The primary goal of this unique device is to perform high-resolution mass separation with small transmission loss since the resolving power is increased while the orbiting ions cool in the ring. Contrary to its linear brother, this circular Paul trap is designed to cool and mass separate the ions over a much longer flight path, thus requiring lower buffer gas pressure. Ions can be accumulated in the ring and extracted either in tangential or perpendicular directions. This way, the trap also serves as a versatile beam distribution device. Design principles are presented and the prototype instrument, under test in Orsay, is described.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.