Abstract

Objectives: We aimed to investigate leaflet kinematics of bioprostheses with a novel high-speed imaging method. Material and methods: High-speed-imaging (1000Hz) was used to evaluate leaflet kinematics of the Carpentier-Edwards Perimount Magna (PM) and Magna Ease (PME) aortic bioprostheses. Both prostheses (diameter 23 mm) were placed inside a model aorta under pulsatile flow conditions. Frequencies (F) and different stroke volumes (S) were simulated. Maximum aortic valve area (AVA), total ejection time (TET), rapid valve opening time (RVOT) and rapid valve closing time (RVCT) as well as opening (OS) and closing (CS) speeds were evaluated. Results: Both bioprostheses showed different results dependent on flow conditions. The test setup was capable of identifying small AVA-differences between both valves (235 vs 202 mm², F60/S60; 272 vs 207 mm²; F70/S80), as well as differences in OS and CS (2.36 vs 1.62 mm²/ms; 2.97 vs 2.44 mm²/ms, F80/S60). TET was comparable (638 vs 645 ms F60/S60; 341 vs 343 ms, F90/S60), while results for RVOT and RVCT were equal, and dependent on frequency and stroke volume. Conclusions: The novel evaluation method is sensitive to detect differences between valves, although differences were found to be small. PM has a larger visible AVA associated with higher opening and closing speeds in contrast to PME.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.