Abstract
Accuracy of videooculography was tested using synthetic eye images. The novel pupil center coordinates extraction method wasproposed. The method exploits averaging of pupil edge coordinates in the same scan line. The novel method accuracy is similar tone ofapproximation by circle method. The proposed method has better computational effectiveness, but is less robust to influence of artifacts.Errors for both methods increase versus noise standard deviation with the same velocity. However the influence of noise is small, ifregion between iris and pupil is sharp. The results of study confirmed that method of videooculography is suitable for recording eyemicromovements. Ill. 6, bibl. 3 (in English, summaries in Lithuanian, English, Russian).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.