Abstract

The dose distribution from a 32P source has been measured and calculated in order to evaluate its application in endovascular irradiation. The source dimension was 27 mm in length and 0.3 mm in diameter and was embedded in the end of a Ni-Ti wire. Dose measurements were performed using radiochromic film in several specially designed tissue equivalent phantoms. Loevinger's point dose kernel was used for the calculation. The approximate dose rate at a radial distance of 1.5 mm from the center of the source was found to be 6.75 cGy/s per GBq (0.25 cGy/s per mCi), which allows the delivery of a therapeutic dose in a short time interval with a satisfactory homogeneity without stepping the source. However, the dose rate falls off almost exponentially along the radial distance. Therefore it may not be suitable for treating large diameter vessel from a centrally located source. The effect of a curved 32P wire source on the radial dose distribution was also investigated. The results showed that for a maximum bend of 180 degrees the dose rate was increased by as much as 20% along the inner radial distance but decreased by as much as 20% along the outer radial distance compared to the dose along a straight wire. However, for curvatures normally encountered in a clinical situation, the dose rate was changed less than 5%.