Abstract

Introduction: The use of cardiopulmonary bypass (CBP; also known as a heart-lung machine) in newborns with complex congenital heart defects may result in brain damage. Magnetic resonance imaging (MRI) assessments cannot be performed safely because the metal components used to construct CBP devices may elicit adverse effects on patients when they are placed in a magnetic field. Thus, this project aimed to develop a prototype MR-conditional circulatory support system that could be used to perform cerebral perfusion studies in animal models. Methods: The circulatory support device includes a roller pump with two rollers. The ferromagnetic and most of the metal components of the roller pump were modified or replaced, and the drive was exchanged by an air-pressure motor. All materials used to develop the prototype device were tested in the magnetic field according to the American Society for Testing and Materials (ASTM) Standard F2503-13. The technical performance parameters, including runtime/durability as well as achievable speed and pulsation behavior, were evaluated and compared to standard requirements. The behavior of the prototype device was compared with a commercially available pump. Results: The MRI-conditional pump system produced no image artifacts and could be safely operated in the presence of the magnetic field. The system exhibited minor performance-related differences when compared to a standard CPB pump; feature testing revealed that the prototype meets the requirements (i.e., operability, controllability, and flow range) needed to proceed with the planned animal studies. Conclusion: This MR-conditional prototype is suitable to perform an open-heart surgery in an animal model to assess brain perfusion in an MR environment.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call