Abstract
The fundamental limit for NMR imaging is set by an intrinsic signal-to-noise ratio (SNR) for a particular combination of rf antenna and imaging subjects. The intrinsic SNR is the signal from a small volume of material in the sample competing with electrical noise from thermally generated, random noise currents in the sample. The intrinsic SNR has been measured for a number of antenna-body section combinations at several different values of the static magnetic field and is proportional to B0. We have applied the intrinsic and system SNR to predict image SNR and have found satisfactory agreement with measurements on images. The relationship between SNR and pixel size is quite different in NMR than it is with imaging modalities using ionizing radiation, and indicates that the initial choice of pixel size is crucial in NMR. The analog of "contrast-detail-dose" plots for ionizing radiation imaging modalities is the "contrast-detail-time" plot in NMR, which should prove useful in choosing a suitable pixel array to visualize a particular anatomical detail for a given NMR receiving antenna.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.