Abstract
The aim of this multicenter study was to assess the feasibility, safety, and accuracy of the intracavitary ECG method for real-time positioning of the tip of different types of central venous catheters. A total of 1444 catheter insertions in adult patients were studied in eight Italian centers (539 ports, 245 PICCs, 325 tunneled CVCs, 335 non-tunneled CVCs). Patients with no visible P wave at the standard baseline ECG were excluded. Depending on the type of catheter and its purpose, the target was to position the tip either (a) at the cavo-atrial junction, or (b) in the lower third of the superior vena cava, or (c) in the upper part of the atrium. The final position was verified by a post-procedural chest x-ray. The method was feasible in 99.3% of all cases. There were no complications potentially related to the method itself. At the final x-ray control, 83% of all tips were positioned exactly at the target; 12.4% were positioned within 1-2 cm from the target, but still in a correct central position; only 3.8% were malpositioned. The mismatch between intra-procedural ECG method and post-procedural x-ray was significantly lower when the x-ray was taken in supine position. Our multicenter study confirms that the intracavitary ECG method for real time verification of tip position is accurate, safe, feasible in all adult patients and applicable to any type of short-term or long-term central venous access device.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.