Abstract
It has been demonstrated that a relatively low-cost resistive NMR imager can be installed in a normal hospital environment without many major or expensive modifications. The magnet can be adjusted to give adequate uniformity and there is sufficient RF shielding to give good quality clinical images (Fig. 3). The fringe field of the magnet of this system, which operates at the lowest field strength of any commercial NMR imager, does not present a problem to imaging unit staff (NRPB, 1984). The long term reliability and detailed specifications with regard to image quality have yet to be determined. These will be determined whilst the imager is being used for clinical studies as part of the national clinical evaluation of NMR imaging supported by the Medical Research Council.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.