Abstract

To test the novel hypotheses that (1) endograft profile can be reduced if the components do not overlap in the carrier catheter and (2) the inside-out deployment technique can produce a device with overlapping layers. This study evaluated an endograft in which the graft and stents were connected to each other in an end-to-end fashion without overlap. The endograft was released into a water flow model to become overlapping layers of graft and stents. Then, a Zenith endograft was modified and deployed using this technique, which reduced the Zenith endograft introducer from 20 to 14 French. Inside-out endograft deployment combined with the end-to-end connection of components without overlap may be an effective method to decrease the entry profile of aortic endografts.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.