Abstract
‡‡ † † , † ‡ We evaluated the influence of volume effect on the measurement of IMRT dose distribution by comparing a 2D-array ion chamber and other dosimeters. Matrix phantom which is a 2D-array ion chamber having volume effect was compared with beam image system and film for the mea surement of dose distribution. Five intensity-modulated radiation therapy plans were created using five fields in thevirtual phantom. The measured dose distribution was compared with the calculated one by radiation treatment planning system and analysis program. We evaluated the conformity of dose distribution by calculating correlation coefficients and gamma values. The highest error rate of 1.3% was associated with matrix phantom in which volume effect in small field sizes was substantial.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.