Abstract

The objective of this study is to evaluate the influence of the diameter of Mini-Implants through bending force tests, taking into account primary stability after one and two insertions. Here, 40 Ti6AI4V alloy Mini-Implants of two different brands and diameters were divided into eight groups, half of which received one insertion in the artificial bone, and the rest received two. All were subjected to a constant bending force using an INSTRON-Electropuls E10000LT (Norwood, MA, USA) until fracture. The smaller-diameter Mini-Implants were less resistant to fracture, but both were able to withstand the necessary loads produced by orthodontic movements. As for the inserts, there were no statistically significant differences. There is an advantage to using 1.6 mm Mini-Implants over 2.0 mm ones, as a smaller diameter does not lead to fracture due to the forces used in orthodontic treatment. Having one or two inserts did not have a statistically significant effect.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.