Abstract
Objective. To investigate the effect of surgical malalignment on contact pressures of fixed and mobile bearing knee prostheses. Design. An experimental set-up was used to measure contact pressure on the tibial component of fixed and mobile bearing knee prostheses subjected to a compression load and surgical malalignment situations were simulated. Background. Mobile bearing knee prostheses were designed to decrease tibiofemoral contact pressure by providing both high congruity and mobility. It was also assumed to accommodate surgical malalignment. However, few studies have reported the effect of malalignment of the tibiofemoral joint on contact pressure of fixed and mobile bearing knee prostheses. Methods. Surgical malalignment situations were simulated to evaluate contact characteristics of tibial component of fixed and mobile bearing knee prostheses. The simulated malalignment conditions include the medial–lateral translation (0.5 and 1 mm), anterior–posterior translation (2 and 4 mm) and internal–external rotation (1°, 3°, 5° and 10°) of the femoral component relative to the tibial component. Fuji pressure sensitive film was used to measure the contact pressure. Results. The greatest increase of maximum contact pressure in the anterior–posterior maltranslation was 7.63% and 7.62% relative to the neutral contact situation in the fixed and mobile bearing designs respectively. In the medial–lateral maltranslation, there was 23.3% in the fixed bearing design and was 22.0% in the mobile bearing design. In the internal/external malrotation, the greatest increase of maximum contact pressure in the fixed bearing design was 27.1%, which was much higher than the mobile bearing design (22.4%). Conclusions. The mobile bearing design can reduce maximum contact pressure more significantly than the fixed bearing design when malalignment conditions of the tibiofemoral joint occurs, especially in the internal/external malrotation. The mobile bearing design offers the advantage of self-adjusting over the fixed bearing design to accommodate surgical malalignment. Relevance This study revealed that the mobile bearing design has smaller maximum contact pressures than the fixed bearing design in knee prosthesis under malalignment biomechanical tests. This result indicates that there is an advantage for a mobile bearing design over a fixed bearing design to accommodate malalignment conditions caused by surgical technique or soft tissues imbalance in total knee arthroplasty.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.