Abstract

Sterilization method changes to the Low Contact Stress (LCS) Total Knee System (DePuy Orthopaedics Inc, Warsaw, Ind) have resulted in progressive improvements in wear resistance. Initially, the system used the traditional gamma in air irradiation sterilization method, but was transitioned to Gas plasma sterilization. The Gas plasma technique did not introduce free radicals that subsequently oxidize with extended shelf aging found with gamma in air irradiation. This oxidation has demonstrated a significant reduction in mechanical properties leading to increased wear. The introduction of Gas plasma was also accompanied by a transition to polyethylene resins that no longer contained calcium stearate, surmised to influence the rate of oxidation. More recently, gamma irradiation and storage of the polyethylene components in an essentially oxygen-free environment imparted by gamma irradiation in a vacuum foil pouch (GVF), has been shown to protect the implant from oxidation. Additionally, the GVF process imparts the beneficial effects of moderate crosslinking that has been demonstrated to improve wear resistance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.