Abstract

BackgroundSome electrical impedance tomography (EIT) devices equip reference electrodes. In practice, it is not uncommon to observe high contact impedance for the reference electrode. The influence of bad contact reference electrode on data quality is unknown. The study aimed to investigate the influence of reference electrode on EIT image reconstruction. MethodsThirty lung healthy volunteers were prospectively examined with EIT. The subjects were spontaneously breathing in supine position. Three scenarios were constructed: 1. Normal measurement; 2. Reference electrode disconnected without recalibration; 3. Reference electrode disconnected, and the measurement restarted after recalibration of the system. EIT-based parameters measuring spatial and temporal ventilation distributions were calculated and compared. A so-call deviation score was calculated to assess the differences in EIT parameters between scenarios 2 and 1, between 3 and 1. ResultsThe absolute differences for all parameters were significantly higher than zero (p < 0.01 for all parameters and scenarios). Deviation score for scenario 2 was 4.5 ± 3.5. Four subjects had a deviation score of 0 in scenario 2 and five subjects had a score of 1. The deviation in scenario 3 was higher (6.1 ± 3.1). No subjects had a score of 0 and only two subjects had a score of 1. ConclusionsFor EIT systems that equips with reference electrode, it is important to ensure the proper contact and functionality of the reference electrode. The EIT data quality would remain unchanged in only a small portion of subjects.