Abstract

The purpose of this study was to clarify the influence of posture change on relative body fat in the bioelectrical impedance analysis (BIA) method. The subjects were 30 Japanese healthy young adult males (age: 19.8 +/- 1.4 years, height: 172.3 +/- 5.8 cm, weight: 67.1 +/- 8.2 kg). We used devices with different body segment inductions, between the hand and foot (H-F BIA) and between hands (H-H BIA), and set four measurement conditions differing in posture (supine or sitting), during rest and measurement. The reliabilities of %BF in the H-H and H-F BIA methods were very high (r = 0.995, 0.966), and the relationship in %BF between the UW method and each BIA method was mid-range (r = 0.767, 0.709). Although there were no differences in %BF among different measurement postures in the H-F BIA method, %BF in the H-H BIA method increased significantly when the posture was changed just before measurement. This indicated that it is necessary to pay attention to the posture change just before measurement in the H-H BIA method.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.