Abstract
The accuracy of atom gravimeters is directly related to the Raman laser, which is used to manipulate the atomic wave packet, and the frequency of the Raman laser could be affected by temperature when the laser polarization is not along the preferred axis of the electro-optic crystal employed in the modulation transfer spectrum (MTS). This effect has been researched by modulating the laser polarization in the MTS in this work. The experimental results show that both the laser frequency and gravity measurement results have a sinusoidal dependence on temperature, and the period of the fluctuation is 0.8 °C. The systematic effect can reach 12.4 μGal when the polarization misalignment is 15°, which is a remarkable contribution to the absolute gravity measurement. The amplitude of this effect could be reduced by adjusting the laser polarization to the crystal's preferred axis. According to the result, the included angle between the laser polarization and the crystal's preferred axis should be smaller than 5° if 2 μGal accuracy is required.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.