Abstract

Magnesium ions can shorten the tissue factor-induced coagulation time. Some blood collection systems with sodium citrate are contaminated with variable amounts of magnesium and influence the results of the prothrombin time (PT) test. The aim of the study was to determine the dose-response relationship between exogenous magnesium chloride added to blood and the PT and the international normalized ratio (INR). Blood specimens from twenty patients on oral anticoagulant therapy were investigated. Four different types of thromboplastin reagents were used: recombinant human, human placenta, rabbit brain, and bovine brain combined with adsorbed bovine plasma. With all four reagents, exogenous magnesium induced a reduction of the apparent INR. Bovine thromboplastin was not as responsive to magnesium as the human and rabbit reagents. The magnitude of the INR deviation induced by 0.1 mmol/l magnesium in the blood was smaller than 10% in all patient samples. At 0.5 mmol/l magnesium in the blood, 10-35% of the patient samples had INR deviations greater than 10%, depending on the thromboplastin reagent used.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.