Abstract

High energy electron beam irradiation of Poly (ether-block-amide) (Pebax) can induce modifications and/or degradation to transpire in the material subsequent to treatment. To minimise this, Pebax was blended with three stabilisers where each formulation was subjected to electron beam radiation at doses of 25, 50 and 75kGy. Mechanical testing revealed that the virgin Pebax and the Pebax blended with Irganox B215 provided the best radiation resistance in terms of the tensile strength, elongation at break and Young's modulus. Upon increase in radiation dose from 25 to 75kGy, a gradual diminution was observed for the melt flow index (MFI) of the virgin Pebax, whereas Pebax blended with Irganox B215 had a minute effect on the properties post irradiation. This study provides evidence that the stabilisers used can either promote undesirable effects or enhance the radiation resistance of Pebax material following radiation exposure.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.