Abstract

Background and purposeIn proton therapy, a 3.5% margin is often used to account for proton range uncertainties, of which computed tomography (CT) image noise is assumed to contribute 0.5%. This work evaluates the noise-sensitivity of three dual-energy computed tomography (DECT)-based methods for mapping proton stopping power relative to water (SPR): Näsmark & Andersson (N&A), Landry-Saito (L-S), and the commercial application DirectSPR. Methods and materialsDECT image data of a CIRS-062M phantom was acquired with CT scanners from two different vendors. Acquisitions were repeated 30 times to account for intra- and inter-scan variations. SPR maps were generated with the three DECT-based methods and range simulated in a commercial treatment planning system. ResultsNoise in input data was amplified in L-S SPR maps, kept level with DirectSPR, while N&A compressed noise overall but displayed sensitivity to the choice of input data, potentially leading to increased noise levels. In our simulations, only N&A improved upon the assumed 0.5% noise contribution to range uncertainty on one scanner. On the other scanner, uncertainties exceeded 0.5% for all three methods. Mitigation of this issue was demonstrated by using a method employing virtual mono-energetic images as input. Increasing imaging radiation dose, as expected, alleviates the problem, while applying noise reduction only helped to a lesser extent. ConclusionsWhile range uncertainty due to noise is small compared to other contributions, it becomes more important as we move towards smaller treatment margins and the noise-sensitivity of SPR mapping methods should be carefully estimated and considered before clinical implementation.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call