Abstract

The Particle Measuring System's 2DC optical array probe has been used as a primary instrument to measure ice particle concentrations since its introduction in 1976. In certain mixed-phase clouds containing cloud droplets in size ranges 30–100 μm, the 2DC has been found to respond to both liquid and ice particles. This is of real concern, since ice crystals occur naturally in tens per liter, while supercooled droplets in this size range can occur in tens to hundreds per liter. In this note, the effect of cloud droplets on the measurement of ice particle concentrations with the 2DC probe is discussed. An example is presented where supercooled droplets overwhelm the probe, making accurate measurements of ice particle concentrations difficult to obtain.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.