Abstract

We measured the peak voltage induced in a sensing loop by a Magstim 200 magnetic stimulator. Coil output varied little for repeated stimulation at the same intensity over a wide range of coil output. In contrast, the first stimulus immediately after a change in intensity was of larger amplitude and showed greater variability than subsequent stimuli. The effect was seen for changes in intensity of 5% and 60% and was greater for reductions than for increases in stimulation intensity. Stimulation immediately after a reduction in intensity from 100% to 40% resulted in peak induced voltages as high as those recorded for repeated stimulation at 43% of maximum coil output. Increased coil output following changes in stimulation intensity may affect measurement of the threshold old for motor cortical stimulation. The effect of a change in intensity could be minimized by delaying stimulation for at least 30 s or discarding the first stimulus after the change.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.