Abstract

Oxygen saturation (SpO2) was measured with a pulse oximeter in ten healthy, young men breathing air. A pulse oximeter probe was attached to the second toe and a laser Doppler probe to the first toe of the same foot for measurement of changes in peripheral blood flow. The pulse oximeter and laser Doppler readings were simultaneously compared when the foot was positioned 40 cm (position 1) above heart level, elevated 10 cm (position 2) above heart level and horizontally at heart level (position 3). Using this experimental human model, we achieved various blood flows. The AC and DC optical signals used for determination of oxygen saturation were recorded from the pulse oximeter and analysed. There was a significant increase (P less than 0.05) between position 1 and 3 in blood flow as measured by the laser Doppler flow meter. The corresponding pulse oximeter readings of haemoglobin saturation also increased significantly (P less than 0.05) comparing these two leg positions. Analysing the AC- and DC optical signals, the AC value of infrared light increased considerably, while the AC value of the red light decreased slightly. The DC values of red and infrared light did not change significantly. In summary, when blood flow was decreased, the ratio of red to infrared transmitted light was changed, resulting in a low SpO2 reading.