Abstract

To find out the effect of central corneal thickness (CCT) and radius of the corneal curvature on intraocular pressure (IOP) measurements using rebound tonometer (RBT) and Tono-Pen in healthy schoolchildren. IOP was measured with Tono-Pen and RBT, respectively, in 165 healthy schoolchildren with a mean age 9.8+/-3.1 (range: 7 to 12 y) years. Corneal radius of curvature (in mm) was determined using a keratometer before CCT and IOP measurements. CCT was measured using an ultrasonic pachymeter after all IOP determinations had been made. The effect of CCT, radius of the corneal curvature, and sex on measured IOP was evaluated by linear regression analysis. The mean IOP readings were 17.47+/-2.7 mm Hg using Tono-Pen, and 16.81+/-3.1 mm Hg using RBT. Tono-Pen measured IOP values slightly greater than that of RBT (P=0.006). Mean CCT was found to be 561.37+/-33 microm. A significant association between measured IOP and CCT was found with each device (r=0.220 for the Tono-Pen, r=0.373 for the RBT; P=0.006 for the Tono-Pen and P<0.0001 for the RBT). The IOP increased 2.2 and 3.7 mm Hg for every 100-microm increase in CCT for the Tono-Pen and the RBT, respectively. The relation between IOP and CCT was not different for boys and girls. Mean radius of the corneal curvature readings was 7.68+/-0.41 mm (42.75+/-1.37 D) for both sexes. There was no significant relationship between either the mean corneal curvature readings, or CCT and IOP (r=0.02; P=0.4 for CCT and r=0.01; P=0.5 for IOP). Both the Tono-Pen and RBT have a systematic error in IOP readings caused by its dependence on CCT. The CCT measurements should be considered to ensure proper interpretation of IOP measurements in children, like in adults. The corneal radius of curvature had no significant effect on measured IOP with each device.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call