Abstract

The aim of this study was to evaluate the influence of two commonly used anticoagulants (K 3EDTA and lithium heparin) on refractometric and spectrophotometric measurement of total protein (TP) concentration in equine peritoneal fluid samples. The influence of a commercial solution of K 3EDTA, a solution of K 3EDTA in distilled water and lithium heparin on the refractometric and spectrophotometric (biuret) quantification of TP content in peritoneal fluid samples was assessed. Total protein concentration measured by refractometry was consistently overestimated in samples with commercial K 3EDTA. The solution of K 3EDTA in distilled water only caused TP overestimation at high K 3EDTA concentrations (>5 μmol/ml). By contrast, lithium heparin did not influence the refractometric values of TP. Neither anticoagulant modified TP values when measured by the biuret method. In conclusion, the use of K 3EDTA as anticoagulant may result in a significant overestimation of TP values of peritoneal fluid samples measured by refractometry.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.