Abstract

One of the primary reasons for utilizing modular tibial polyethylene inserts (MTPI) at the time of total knee arthroplasty is to have the ability to simply exchange the polyethylene at the time of revision surgery when polyethylene failure has occurred. During a 2-year period from January 1993 to December 1994, 62 revision total knee arthroplasties were reviewed from five different institutions in North America, which were performed secondary to modular tibial insert failure. In 55 cases (88.7%), significant scoring and/or damage to the femoral and/or tibial components occurred necessitating revision of one or both components. This series does not support the premise that polyethylene exchange is common at the time of revision surgery for MTPI failure. Of the patients 88.7% had MTPI failure resulted in femoral and/or tibial component revision secondary to surface damage to the femoral component and/or tibial component baseplate. In order to avoid femoral and tibial component wear secondary to MTPI failure, early recognition is necessary to allow the much simpler operation consisting of polyethylene liner exchange only.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.