Abstract
We have studied the effect of brain tissue (gray matter-white matter) heterogeneity and computerized tomography (CT) noise on the accuracy of xenon CT measurements of regional cerebral blood flow (rCBF) based upon "autoradiographic" and multiple-scan washin protocols. The results of our mathematical analysis indicate that both protocols are associated with a variety of measurement errors that lead to unacceptable and, to a large extent, unpredictable uncertainties in calculated values of rCBF. Brain tissue heterogeneity and high volumetric flow rates may--even in the absence of movement artifact and CT noise--lead to measurement errors in excess of 20%. Moreover, CT noise is additive in regard to these errors, and constitutes the most confounding variable of all.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of cerebral blood flow and metabolism : official journal of the International Society of Cerebral Blood Flow and Metabolism
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.