Abstract
We report the in vitro longevity of a conventional soda lime carbon dioxide absorbent and an absorbent free from strong alkali (Amsorb). Although the times taken to breakthrough of carbon dioxide (> 0.5%) within an in vitro low flow breathing system were shorter with the alkali-free absorbent, we found that the size and shape of the absorbent container was the major factor in determining the efficiency of the carbon dioxide absorbents.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.