Abstract

Ambient dose equivalent dosemeters are widely distributed in nuclear facilities for routine continuous monitoring or nuclear accident emergency monitoring, they are inconvenient to be disassembled and sent to the metrology laboratory for calibration. To ensure the accuracy of such dosemeters, the research of in-situ calibration method has been carried out. Ambient dose equivalent secondary standard ionization chamber and portable gamma ray irradiation facility have been designed for in-situ calibration. The experiments of in-situ calibration were carried out in five typical sites located in China Institute of Atomic Energy, results showed that the relative deviation between in-situ calibration factors and metrology laboratory calibration factors are within 5%.The combined standard uncertainty of the in-situ calibration factor is 2.8%, and the expanded uncertainty is 5.6% (k=2).The in-situ calibration method can meet the calibration requirements of the fixed ambient dose equivalent dosemeters.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.