Abstract

Optimization of the operating parameters of U-shpaed DC arc plasma and spectrometer parameters has been undertaken to explore the possibilities of improving its detection power. It is demonstrated, with a U-shaped arc as an example, that the limits of detection, in addition to well-defined parameters as described by Boumans1,2 and Winge,3 depend on the signal integration time. It is shown that with increasing integration time, the limits of detection are decreased within some limits and that the precision and concentration sensitivity are improved as well. A mathematical expression for the dependence of the detection limit on the integration time is presented. To increase the reliability of the measurement of the mentioned parameters, the working conditions were optimized for the following analytes: Ag, Al, Au, Cr, Fe, Mn, Ni, Pb, Pd, Pt, and V. The obtained limits of detection are comparable or better than those obtained by ICP for the elements studied. It was estimed that the possibility exists for their further improvement up to 10 times.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.