Abstract

Two non-invasive anal plug electrodes of similar size have been compared, one with the electrode plates orientated circularly in the anal canal and the other with the plates in the long axis of the anal canal. There was a significant increase in the amplitude in the EMG signals recorded at rest and during squeeze from the external anal sphincter with a longitudinally placed electrode in 117 patients. Inappropriate contraction of the external anal sphincter when straining at stool was more readily detected using the longitudinal electrode in 52 patients investigated for intractable constipation. The longitudinal electrode detected the amplitude of the response to the elicitation of a pudeno-anal reflex more readily than the circular electrode. When in 12 of the 117 the pudeno-anal reflex EMG signal was either absent or not detected with the circumferential plug electrode, the longitudinal electrode detected the presence of a low amplitude response in 11 of these. When the non-invasive longitudinal electrode was compared to invasive fine wire stainless steel electrodes, a correlation was found for external anal sphincter resting EMG (r = 0.99, p less than 0.01), voluntary squeeze EMG (r = 0.99, p less than 0.001) and strain EMG (r = 0.91, p less than 0.01). The longitudinal anal plug electrode thus facilitates surface acquisition of EMG activity.