Abstract
BackgroundWith the advent of the new high-sensitivity troponin assays, it is becoming critical to measure troponin accurately to low concentrations. To ensure assay performance is acceptable, appropriate QC must be run. MethodsIn addition to the routine use of commercial QC material, we prepared pools of human QC material with low troponin concentrations close to the limit of quantitation, and ran these regularly on our laboratory analysers. ResultsOver 3 years we found no drift or shift in our hs-cTnI assay. We found that only the very low concentration human QC material gave warning of precision problems with the hs-cTnI assay. At the time of the documented poor assay precision, the higher concentration QC material indicated satisfactory performance. ConclusionsChoice of QC material with an appropriate concentration is important for any assay. For hs-cTn assays, it is of particular importance to use control material with a concentration near to the limit of quantitation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.