Abstract

Today, the increase in demand for electrical energy has revealed a more reliable and higher quality energy concept. This subject has taken its place in electrical engineering as Power Quality or Electric Energy Quality and has made a point for its importance with the studies on it. Especially in the process of working with devices containing electronic circuits with direct current and using the alternating current offered from the network by using rectifiers, the devices may fail or become disabled due to self-protection circuits in case of power quality problems arising from the mains or the rectifiers of the device. In this article, a detailed analysis is carried out considering the problem may be due to the power quality due to the fact that the MRI device in Kirikkale University Faculty of Medicine is in continuous failure mode. In addition, the analyzes made in accordance with TS EN 50160 standard within the scope of the research were analyzed in computer environment and the results were shared in our article. In this article, the relationship between power quality problems and grounding is also examined.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.