Abstract

ECG distortion, mimicking changes seen in acute myocardial infarction, occurred in traces transmitted by the South East Coast NHS Ambulance Trust. This was due to low frequency phase nonlinearity produced by baseline filtering. The very narrow calibration pulses made the detection of this distortion difficult. A 1 mV square wave signal revealed the cause. Adequate calibration pulses 0.3 s wide should be recorded with all ECGs, to show that the signal is being reproduced reliably.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.