Abstract
Indirect dosimetry to calculate peak skin dose (PSD) is generally based on reference air kerma or kerma-area-product, with effects of table attenuation, inverse square law and backscatter factor applied. When the incident x-ray beam angle changes, these factors would change as well. The purpose of this study is to identify the impact of incident x-ray beam angle on the accuracy of indirect PSD calculation and develop a correction method. Monte Carlo simulation was conducted to assist analytical equation derivation and to validate the developed analytical method. Direct PSD measurements were performed a Siemens Artis Zee biplane fluoroscopy and a GE OEC C-Arm at different angles to validate the proposed correction method and its independence of fluoroscopy systems. Compared with MC simulated PSD, the derived analytical equation successfully predicts these PSD changes for incident angles of 0, 15, 30 and 40 degrees, with the residual error magnitude of 0.3%–3.1% corresponding to different SSDs. The agreement between calculated PSD also agrees well with direct measurement.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.