Abstract

Phospholipids (PLs) are a component of cellmembranes, biological fluids and tissues. These compounds are problematic for the bioanalytical chemist, especially when PLs are not the analytes of interest. PL interference with bioanalysis highly impacts reverse-phase chromatographic methods coupled with mass spectrometric detection. Phospholipids are strongly retained on hydrophobic columns, and can cause significant ionization suppression in the mass spectrometer, as they outcompete analyte molecules for ionization. Strategies for improving analyte detection in the presence of PLs are reviewed, including in-analysis modifications and sample preparation strategies. Removal of interfering PLs prior to analysis seems to be most effective atmoderating thematrix effects fromthese endogenous cellular components, and has the potential to simplify chromatography and improve column lifetime. Products targeted at PL removal for sample pre-treatment, as well as products that combine multiplemodes of sample preparation (i.e. Hybrid SPE), show significant promise inmediating the effect on PL interference in bioanalysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.