Abstract

The Multiple Coulomb Scattering (MCS) model plays an important role in accurate MC simulation, especially for small field applications. The Rossi model is used in MCNPX 2.7.0, and the Lewis model in Geant4.9.6.p02. These two models may generate very different angular and spatial distributions in small field proton dosimetry. Beside angular and spatial distributions, step size is also an important issue that causes path length effects. The Energy Fraction (EFAC) value can be used in MCNPX 2.7.0 to control step sizes of MCS. In this study, we use MCNPX 2.7.0, Geant4.9.6.p02, and one pencil beam algorithm to evaluate the effect of dose deposition because of different MCS models and different EFAC values in proton disequilibrium situation. Different MCS models agree well with each other under a proton equilibrium situation. Under proton disequilibrium situations, the MCNPX and Geant4 results, however, show a significant deviation (up to 43%). In addition, the path length effects are more significant when EFAC is equal to 0.917 and 0.94 in small field proton dosimetry, and using a 0.97 EFAC value is the best for both accuracy and efficiency

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.