Abstract

PurposeIn this study, we aimed at investigating the impact of deterministic signal loss on image quality and, thus, on optical coherence tomography angiography (OCTA) measurements performed by the RTVue-XR Avanti System.MethodsAbsorptive filters with different optical densities (ODs) were used to simulate signal loss in a controlled way in 30 eyes from 15 healthy subjects. Scan Quality (SQ), provided by the AngioVue software, was applied as a measure of image quality.ResultsAssessing the effect of decreased light transmittance on SQ values as well as that of attenuated image quality on OCTA parameters revealed a strong systematic dependence between these quantities. Attenuated image quality was associated with significantly decreased macular and peripapillary vessel density (VD) values, and we calculated a correction factor of 2.27% to 3.97% for a one-unit change in SQ for the different VD parameters.ConclusionsOur findings suggest that the influence of systematic changes in image quality on OCTA parameters needs to be considered during patient follow-up in order to make valid assessment of progression.Translational RelevanceFor accurate evaluation of longitudinal changes in OCTA parameters, equal scan quality or using a correction factor is suggested.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.