Abstract
The most commonly used method to deliver aerosolized drugs to the lung is with pressurized metered-dose inhalers (pMDIs). The spray actuator is a critical component of pMDI, since it controls the atomization process by forming aerosol plumes and determining droplet size distribution. Through computational fluid dynamics (CFD) simulations, this study investigated the effect of two different nozzle types (single conventional and twin nozzles) on drug deposition in the mouth-throat (MT) region. We also studied the behavior of aerosol plumes in both an open-air environment and the MT geometry. Our study revealed that spray aerosol generated in an unconfined, open-air environment with no airflow behaves distinctly from spray introduced into the MT geometry in the presence of airflow. In addition, the actuator structure significantly impacts the device's efficacy. In the real MTmodel, we found that the twin nozzle increases drug deposition in the MT region, and its higher aerosol velocity negatively affects its efficiency.
Submitted Version (Free)
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.