Abstract

The design and performance of the Hygroscopic Condenser Humidifier (HCH) are described. In principle the HCH consists of two parts, a conventional Heat-Moisture-Exchanger (HME) and a hygroscopic unit. The hygroscopic action is shown to improve the water retention efficiency of the device by about a factor of two as compared with optimal HME designs. As a result, humidification levels corresponding to around 80% relative humidity at 37 degrees C are obtained in the trachea and this is also achieved when completely dry gases are delivered to the patient. The unit can therefore be used for all procedures in anaesthesia and in intensive care.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.