Abstract

Using the basic spectrometer trajectory equation for motion in an ideal 1 / r potential derived in Eq. (101) of part I [T.J.M. Zouros, E.P. Benis, J. Electron Spectrosc. Relat. Phenom. 125 (2002) 221], the operational characteristics of a hemispherical deflector analyser (HDA) such as dispersion, energy resolution, energy calibration, input lens magnification and energy acceptance window are investigated from first principles. These characteristics are studied as a function of the entry point R 0 and the nominal value of the potential V ( R 0 ) at entry. Electron-optics simulations and actual laboratory measurements are compared to our theoretical results for an ideal biased paracentric HDA using a four-element zoom lens and a two-dimensional position sensitive detector (2D-PSD). These results should be of particular interest to users of modern HDAs utilizing a PSD.